32
ICS Aircal
Notes on EMC (Electromagnetic Compatibility)
Guidance and manufacturer's declaration - electromagnetic immunity - for equipment and systems within Professional Healthcare use
environment
The ICS Aircal device is intended for use in the electromagnetic environment specified below. The user of the ICS Aircal device should ensure that it is
used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
3 V rms
150 kHz to 80 MHz
6 V rms
ISM Bands
3 V rms
150 kHz to 80 MHz
6 V rms
ISM Bands
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
80 MHz to 2.7 GHz
Proximity fields from RF wireless
communications
IEC 61000-4-3
27 V/m
385 MHz
28 V/m
450 MHz
27 V/m
385 MHz
28 V/m
450 MHz
Separation distance between any
electronic parts of the ICS Aircal device
and any RF wireless communication
equipment must be more than 30 cm
(11.8 inches).
9 V/m
710 MHz, 745 MHz, 780
MHz
9 V/m
710 MHz, 745 MHz, 780
MHz
28 V/m
810 MHz, 870 MHz, 930
MHz
28 V/m
810 MHz, 870 MHz, 930
MHz
28 V/m
1720 MHz, 1845 MHz, 1970
MHz
28 V/m
1720 MHz, 1845 MHz, 1970
MHz
28 V/m
2450 MHz
28 V/m
2450 MHz
9 V/m
5240 MHz, 5500 MHz, 5785
MHz
9 V/m
5240 MHz, 5500 MHz, 5785
MHz
Note 1: At 80 MHz and 800 MHz the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
a.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the ICS Aircal device is
used exceeds the applicable RF compliance level above, the ICS Aircal device should be observed to verify normal operation. If abnormal
performance is observed, additional measures might be necessary, such as reorienting or relocating the ICS Aircal device.
b.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.