ICS Aircal
Getting started
Filling
Turn off the ICS Aircal device and unplug it from the electrical wall outlet.
Wipe off any dust on the top of the case.
Facing the front panel, locate the small cover at the back right side. Using the Allen key (supplied) unscrew the
reservoir cover and remove.
A.
Reservoir cover
Using the Allen key (supplied with the cradle), unscrew the internal cover and remove.
Warning
• To prevent electrical shock, the ICS Aircal device must be turned off and disconnected from the
electrical wall outlet when filling or refilling the water reservoir.
Note
•
Use this procedure:
1.
If the ICS Aircal device was just received from shipping and is dry. It must be filled and primed.
2.
If the water level in the display has gone to zero because the user never refilled it when the orange bar was
present.
3.
If the ICS Aircal device was tilted at more than 20° with the water pump running.
If none of these criteria were met and the water level display is showing an orange bar and needs a refill, follow the
instructions in
Note
•
Only
use distilled water. Using common water from the tap will result in mineral deposits and will harm
the ICS Aircal device.