11
ICS Aircal
Getting started
A.
Internal cover
Expand the funnel supplied with the ICS Aircal device and place it in the refill hole of the reservoir.
A.
Funnel
Pour water into the funnel to fill the reservoir.
Monitor the water level closely. Fill to the bottom of the lip. Do NOT overfill the reservoir.
Remove the funnel.
Wipe up any excess water that might have spilled into the small indentions where the internal cover is placed.
Replace the internal cover.
Priming
Plug the power cable of the device into the electrical wall outlet and turn it on.
Locate the priming switch on the back panel .
Press the priming switch to turn on and hold 5 seconds.
Release the priming switch to turn off and wait 5 seconds.
Repeat steps
4
and
5
for a total of 10 times on and 10 times off.
If the temperature reaches the set temperature, then the irrigator is primed. Continue to step
10
.