7
ICS Aircal
Getting started
2.2
Installation and setup
2.2.1
Removing the transportation screws
To prevent possible damage during shipping, the air pump housing is secured with transportation screws. These
screws must be removed to allow the housing to move and thereby reduce the amount of noise the air pump
produces. It is important to keep the transportation screws to secure the housing if the device ever needs to be
shipped.
Remove transportation screws (A) and (B).
Store them with your syringe kit, funnel, and Allen key.
2.2.2
Installing the Delivery Head Cradle
Before installing the cradle for the delivery head, determine how it should be mounted:
•
The cradle can be mounted on the left or right side of the unit.
•
For units that are placed away from the edge of the table, it is recommended that the cradle is mounted in the
horizontal position (A).
For units that are placed at the edge of the table, it is recommended that the cradle is mounted at an angle (B).
Caution
• The ICS Aircal system is intended to be used only by qualified medical personnel.
Note
•
Install the cradle prior to filling the water reservoir. Installation requires that the ICS Aircal device is tilted
more than 20°. If the water reservoir has water in it, tilting the device could introduce air into the system. The risk
of damage to the system is small, but priming of the water reservoir would be required to make the system
operate properly and not overheat.