Warning
•
Do not use this device outside the operating and storage environments described in this manual.
Warning
•
Keep the unit away from liquids. Do not allow moisture inside the unit. Moisture inside the unit can
damage the instrument and it may result in a risk of electrical shock to the user or patient.
Warning
•
Do not use the instrument in the presence of flammable agents (gases) or in an oxygen-rich envir-
onment.
Warning
•
Changes or modifications not expressly approved by the manufacturer could void the user's authority
to operate the equipment.
Warning
•
Computer and printer must be placed out of reach of the patient, i.e. not closer than approx. 1.5
meters/5 ft.
Warning
•
Any PC connected to the device must comply with the requirements of IEC 62368-1:2020.
Warning
•
The use of this device in domestic establishments is allowed, only under the jurisdiction of a health
care professional.
Warning
•
Refer to the Techical Specifications section of this manual for the environments for which the Medical
Electrical System (ME System) is suitable and the Notes on EMC (Electromagnetic Compatibility).
Warning
•
Electrostatic discharge (ESD) may occur through the connectors of HI-PRO 2.
Warning
•
This device contains a Type BF applied part symbol. Refer to the Definition of Symbols section for more
information.
Warning
•
Do not connect this device to other devices that do not comply with the safety and EMC standards
described in this manual.