RF emissions
CISPR11
Group 1
Aurical Aud uses RF energy only for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR11
Class B
Aurical Aud is suitable for use in all environments, including domestic environments and those
directly connected to the public low-voltage power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions IEC
61000-3-2
Complies
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Complies
Guidance and manufacturer's declaration - electromagnetic immunity for all equipment and systems
Aurical Aud is intended for use in the electromagnetic environment specified below. The user of Aurical Aud should ensure that it is used in such an envir-
onment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
+/- 8 kV contact
+/- 2 kV, +/- 4 kV,
+/- 8 kV, +/- 15 kV air
+/- 8 kV contact
+/- 2 kV, +/- 4 kV,
+/- 8 kV, +/- 15 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humid-
ity should be at least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
+/- 2 kV for power supply lines
+/- 1 kV for input/output lines
+/- 2 kV for power supply lines
+/- 1 kV for input/output lines
Mains power quality should be that of a typical com-
mercial or hospital environment.
Surge
IEC 61000-4-5
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
+/- 2 kV DC input line(s) to earth
+/- 1 kV DC input line(s) to line(s)
+/- 2 kV I/O line(s) to earth
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
+/- 2 kV DC input line(s) to earth
+/- 1 kV DC input line(s) to line(s)
+/- 2 kV I/O line(s) to earth
Mains power quality should be that of a typical com-
mercial or hospital environment.
Voltage dips, short inter-
ruptions and voltage vari-
ations on power supply input
lines
IEC 61000-4-11
0 % U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
and 315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30 cycles
Single phase: at 0°
0 % U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
and 315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30 cycles
Single phase: at 0°
Mains power quality should be that of a typical com-
mercial or hospital environment. If the user of the Aur-
ical Aud requires continued operation during power
mains interruptions, it is recommended that the Aurical
Aud be powered from an uninterruptible power supply
or a battery.
Voltage interruptions on
power supply input lines
IEC 61000-4-11
0 % U
T
; 250/300 cycles
0 % U
T
; 250/300 cycles
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
30 A/m
No relevant ports that could be
affected
Power frequency magnetic fields should be at levels char-
acteristic of a typical location in a typical commercial or
hospital environment.
U
T
is the AC mains voltage prior to application of the test level.