Menu item
Icon
Description
Frequency Resolution
(Frequency Res-
olution)
The options for frequency resolutions are 1/6, 1/12, 1/24 and 1/48
octave as well as 1 Hz. Select the different tone stimulus resolutions
from the toolbar or from
Tools
>
Options
>
Audiometry
>
General
(Tools > Options > Audiometry > General).
You can store up to 24 points for each audiometry curve. You will be
prompted if you try to store more than the maximum number of points.
Monitoring
(Mon-
itoring)
Enables or disables the monitor speaker for monitoring stimuli presen-
ted to the patient from the
Stimulus
(Stimulus) or
Masking
(Masking)
channel.
Talk Forward
(Talk
Forward)
Enables communicating with the patient in the sound booth. This will
display the
Talk Forward
(Talk Forward) dialog box, where you can con-
trol the talk forward microphone sensitivity and the output level (in dB
HL) to the patient.
Select Orientation
(Select Orientation)
Click to select the perspective of the patient's ears as presented on the
screen for graph and table views.
You can also select the location of the stimulus control.
11
Proper transducer placement
Headphones
1. Loosen the headband and place both the left and right side of the headphones simultaneously.
Note
•
If the headphones are not placed properly, there is risk of causing the ear canal to collapse which will result in
elevated thresholds.
2. Aim the center of the headphones towards the patient's ear canals and gently place them against the ears.
3. Tighten the headband while holding the headphones in place with your thumbs.
4. Examine the placement of the headphones to make sure they are level, and properly positioned.
Insert Earphones
1. Select the largest foam eartip that will fit into the patient's ear.
If the eartip is too small the sound will leak out and the sound level will not be accurate at the eardrum.
Insert earphones have greater attenuation between ears especially at the low frequencies; this reduces the need for
masking.
2. It is best to clip the insert earphone transducers behind the child or on the back of their clothing and then fit the
foam eartip into the child's ears.
12
Aurical Aud
11 Proper transducer placement