15.2
Cleaning
Warning
•
Clean the device and the accessories using the following instructions.
The device
•
Remove dust using a soft brush.
•
Use a soft, slightly damp cloth with a small amount of mild detergent.
Warning
•
Keep the unit away from liquids. Do not allow moisture inside the unit. Moisture inside the unit can
damage the instrument and it may result in a risk of electrical shock to the user or patient.
Accessories
Caution
•
These parts are in contact with your patients and should be cleaned between uses.
•
Headphones and bone conductors
Use a non-alcohol based wipe (e.g. Audiowipe) to clean the headphones and bone conductor between patients.
•
Eartips for Insert Earphones
The eartips are single use and should be disposed of after use.
Disposal
There are no special requirements for the disposal of eartips, i.e. they can be discarded according to local regulations.
15.3
Calibration
Annual calibration
The audiometer, headphones, bone conductors, and sound field speakers must be calibrated once a year by your author-
ized service department.
Remote calibration
You can order a transducer and get the calibration data installed via remote support. The calibration data is included in
your shipment on a USB memory stick (or supplied by technical support during the installation).
To import calibration data:
1. Connect the new transducer to your audiometer.
2. Connect the audiometer to your Otosuite PC.
3. Insert the USB memory stick in an empty slot on your PC.
4. Call your Natus technical support team. They will use the application TeamViewer to ensure correct remote install-
ation of the new calibration data on your system.
TeamViewer is located at
Help
(Help) >
Remote support
(Remote support).
20
Aurical Aud
15 Maintenance