The technician installs the calibration data via the menu function
Tools
(Tools) >
Audiometer service
(Audiometer ser-
vice). The data is password protected.
5. When the installation has ended, hold the new transducer within hearing distance and cautiously perform a listening
check.
The purpose of the check is to ascertain that the transducer is functioning correctly (without wrong or excessive sound
levels), not to verify the exact calibration.
Note
•
Note that calibration has been performed only on the transducers supplied. If you wish to use any other trans-
ducer for testing with the device, please contact your local distributor first.
16
Other references
For more information, see the online Help in Otosuite, which contains detailed reference information about Aurical Aud
and the Otosuite modules.
For Otosuite installation instructions, see the Otosuite Installation Guide, on the Otosuite installation medium.
For Troubleshooting information, refer to the Aurical Aud Reference Manual.
17
Technical specifications
17.1
Aurical Aud
Type identification
Aurical Aud is type 1081 from Natus Medical Denmark ApS.
Channels
Aurical Aud is a two channel audiometer.
Frequency range
Insert earphones:
125 - 8000 Hz
TDH39 earphones:
125 - 12500 Hz
HDA 300 headphones:
125 - 12500 Hz
ME-70 headphones:
125 - 12500 Hz
HOLMCO headset:
125 - 12500 Hz
Bone conductor (BC):
250 - 8000 Hz
SF:
125 - 12500 Hz
Aurical Aud
21