Chapter 5
System Configuration
© National Instruments Corporation
5-43
VME-MXI-2 User Manual
devices, and finds the VME-MXI-2 in VMEbus Mainframe #4 and
the VME-MXI-2 in VMEbus Mainframe #5.
5.
Enables the logical address window of the VME-MXI-2 in
VMEbus Mainframe #4 for the entire inward mapping range
of 0 to FF. Scans all logical addresses, skipping all previously
encountered devices. Because VMEbus Mainframe #4 has no
VXIbus devices, the logical address window does not need to be
programmed.
6.
Enables the logical address window of the VME-MXI-2 in
VMEbus Mainframe #5 for the entire inward mapping range
of 0 to FF. Scans all logical addresses, skipping all previously
encountered devices and previously defined address ranges.
Because VMEbus Mainframe #5 has no VXIbus devices, the
logical address window does not need to be programmed.
7.
Sets the logical address window of the VME-MXI-2 found at
logical address 61 to cover the ranges of the VME-MXI-2 in
VMEbus Mainframe #4 (62) and the VME-MXI-2 in VMEbus
Mainframe #5 (63). Enables the logical address window of the
VME-MXI-2 at logical address 61 with an outward range of 62 to
63 by writing the value 4762 hex to the Logical Address Window
Register (Base/Size format).
8.
Sets the logical address window of the VME-MXI-2 found in
VMEbus Mainframe #3 at logical address 60 to cover the devices
in that mainframe (60 to 61) and the ranges required of its Level 2
devices: the VME-MXI-2 in VMEbus Mainframe #4 (62) and the
VME-MXI-2 in VMEbus Mainframe #5 (63). Enables the logical
address window of the VME-MXI-2 at logical address 60 with an
inward range of 60 to 63 by writing the value 6660 hex to the
Logical Address Window Register (Base/Size format).
9.
Enables the logical address window of the VXI-MXI-2 in VXIbus
Mainframe #2 for the entire inward mapping range of 0 to FF.
Scans all logical addresses, skipping all previously encountered
devices and defined ranges. Finds the Slot 0 device and uses it to
move all DC devices in VXIbus Mainframe #2 to the lowest
unused logical addresses. No more VXI-MXI-2 interfaces are
found. The RM enables the logical address window for the
VXI-MXI-2 in VXIbus Mainframe #2 with an inward range of
40 to 5F hex by writing the value 6340 hex to the Logical Address
Window Register (Base/Size format).
10. Sets the logical address window of the VME-MXI-2 found in
VMEbus Mainframe #1 at logical address 0 to cover the
devices connected to that extender: the VME-MXI-2 in
VMEbus Mainframe #3 (60 to 63), the VXI-MXI-2 in VXIbus