2-6
|
ni.com
Chapter 2
Digital I/O
Device Routes
tab in MAX. Refer to
Device Routing in MAX
in the
NI-DAQmx Help
or the
LabVIEW Help
for more information.
If the NI 6612 receives a DI Sample Clock when the FIFO is full, it reports an overflow error to
the host software.
You can sample data on the rising or falling edge of DI Sample Clock.
Routing DI Sample Clock to an Output Terminal
You can route DI Sample Clock out to any PFI <0..39> terminal. The PFI circuitry inverts the
polarity of DI Sample Clock before driving the PFI terminal.
Other Timing Requirements
The NI 6612 only acquires data during an acquisition. The device ignores DI Sample Clock
when a measurement acquisition is not in progress. During a measurement acquisition, you can
cause the device to ignore DI Sample Clock using the DI Pause Trigger signal.
The DI timing engine on the device internally generates DI Sample Clock unless you select some
external source. DI Start Trigger starts this timing engine and either software or hardware can
stop it after a finite acquisition completes. When using the DI timing engine, you also can specify
a configurable delay from DI Start Trigger to the first DI Sample Clock pulse.
By default, this delay is set to two ticks of the DI Sample Clock Timebase signal.
Figure 2-4.
DI Sample Clock and DI Start Trigger
DI Sample Clock Timebase Signal
By default, the NI 6612 routes the onboard 100 MHz timebase to DI Sample Clock Timebase.
You can route many signals to DI Sample Clock Timebase. To view the complete list of possible
routes, see the
Device Routes
tab in MAX. Refer to
Device Routing in MAX
in the
NI-DAQmx
Help
or the
LabVIEW Help
for more information.
DI Sample Clock Timebase is not available as an output on the I/O connector. DI Sample Clock
Timebase is divided down to provide one of the possible sources for DI Sample Clock. The
DI
Sa
mple Clock Time
bas
e
DI
S
t
a
rt Trigger
DI
Sa
mple Clock
Del
a
y
from
S
t
a
rt
Trigger
Содержание NI 6612
Страница 1: ...PCIe 6612...
Страница 2: ...NI 6612 User Manual NI 6612 User Manual November 2013 374008B 01...
Страница 11: ...NI 6612 User Manual National Instruments xi Appendix B Technical Support and Professional Services...