© National Instruments
|
5-1
5
Counter Signal Routing and
Clock Generation
NI 6612 has flexible signal routing features. Signals can be routed to or from the following
sources:
•
User input through the PFI terminals
•
Any of the eight counters
•
The digital I/O circuits
•
Other devices in your system through RTSI
•
User input through the PXI_Star terminal
•
Clocking circuitry
Clock Routing
Figure 5-1 shows the clock routing circuitry of an NI 6612 device.
Figure 5-1.
Clock Routing Circuitry
PXI_Trigger <0..7>
On
b
o
a
rd
100 MHz
O
s
cill
a
tor
Extern
a
l
Reference
Clock
(To RT
S
I <0..7>,
PXI_Trigger <0..7>
& PFI <0..
3
9>
O
u
tp
u
t
S
elector
s
)
10 MHz RefClk
PLL
÷
5
÷
200
DAQmx Timing Property
RefClk.
S
rc
PXIe_CLK100
PXI_
S
TAR
100 MHz
Time
bas
e
100 kHz
Time
bas
e
20 MHz
Time
bas
e
PFI
PXIe-D
S
TAR<A, B>
RT
S
I <0..7>
÷
10