© National Instruments
|
3-13
Figure 3-15.
Frequency Measurement: Sample Clock with Averaging
At each sample clock, the device stores
T1
and
T2
in a buffer. DAQmx reads these values, and
uses the known frequency of the timebase (
f
k
), to calculate the frequency of the input signal (
f
x
)
as:
The maximum error and maximum frequency error for this method, are given by:
To use the Sample Clock (with Averaging) method, configure the following:
•
DAQmx Create channel (CI-Frequency)
—Use this VI or function to create the channel.
•
DAQmx Timing (Sample Clock)
—Use this VI or function to set the number of samples,
sample clock source, and other properties.
•
CI.Freq.EnableAveraging
—Set this property to True.
By default, the counter measures the frequency on a default PFI terminal (refer to Chapter 5,
Counter Signal Routing and Clock Generation
, for more information) and use an onboard
100 MHz clock as the timebase. To change the signals used for this measurement, configure the
following:
•
CI.Freq.Term
—The signal-to-measure comes from an input terminal. To change the
signal-to-measure, set this property to a different terminal.
S
ign
a
l to
Me
asu
re (f )
Co
u
nter
Time
bas
e (f )
Sa
mple
Clock (f )
DAQmx
S
t
a
rt T
as
k
T1 T2
1 7
T1 T2
1 7
2 10
T1 T2
1 7
2 10
1 6
B
u
ffer
1
2
1
7
10
6
Sa
mple 2
Sa
mple
3
Sa
mple 1
x
k
s
f
x
f
k
T
1
T
2
------
×
=
Maximum Error (%)
f
x
f
k
f
x
f
s
---
1
–
×
----------------------------
=
100%
×
Maximum Frequency Error (Hz)
f
x
f
x
f
k
f
x
f
s
---
1
–
×
----------------------------
×
=