Appendix: HART Specific Information
MSA
ULTIMA X ® Series
GB
117
Command #179: Write Sensor Data Sheet Reset Control
Writes a data sheet reset command to ULTIMA X Gas Monitor. This command
causes the ULTIMA Monitor to reset the current sensor data sheet to factory default
settings. This command will set certain device warning status bits and require the
user to re-calibrate the sensor. At present, the only valid number for this command
is 1.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
Byte
Format
Description
0
Unsigned
Sensor reset control
Byte
Format
Description
0
Unsigned
Sensor reset control
Code
Class
Description
0
Success
No Command-specific errors
1 - 2
Undefined
3
Error
Parameter too large
4
Error
Parameter too small
5
Error
Too few data bytes
6
Undefined
7
Error
In write protect mode
8 - 15
Undefined
16
Error
Access restricted
17 - 31
Undefined
32
Error
Busy
33 - 127
Undefined
Содержание ULTIMA X-Series
Страница 1: ...Operating Manual ULTIMA X Series Gas Monitors Order No 10046690 09 ...
Страница 2: ...MSA AUER GmbH Thiemannstrasse 1 D 12059 Berlin Germany MSA AUER GmbH All rights reserved ...
Страница 28: ...MSA Description ULTIMA X Series 28 GB Fig 5 ULTIMA XA Fig 4 ULTIMA Remote sensor module non reactive gas ...
Страница 69: ...Appendix Electrical Installation MSA ULTIMA X Series GB 69 9 3 Installation Outline Drawing CE ULTIMA XA ...
Страница 72: ...MSA Appendix Electrical Installation ULTIMA X Series 72 GB 9 6 HART module M6 109 45 4 309 ...
Страница 158: ...MSA Appendix HART Specific Information ULTIMA X Series 158 GB Sensor Trim Point Screen Fig 52 Sensor trim point screen ...