Appendix: HART Specific Information
MSA
ULTIMA X ® Series
GB
107
Command #134: Read Last Cal Date
Returns the ULTIMA last calibration date of the currently connected sensor.
Request Data Bytes
None.
Response Data Bytes
Command #135: Read Gas Table
This command returns the ULTIMA sensor Gas Table currently in use. The Gas Ta-
bles are linearisation reference tables used with certain sensors to provide accurate
response for different gases from the same sensor.
Request Data Bytes
None.
Response Data Bytes
Command #136: Read Input Voltage Value
Returns the ULTIMA input supply voltage value. This number should be in the
range of 8-30 volts DC.
Request Data Bytes
None.
Response Data Bytes
Byte
Format
Description
0-2
Unsigned
Last sensor calibration date
Byte
Format
Description
0
Unsigned
Gas Table Number [
Byte
Format
Description
0-3
Float
Input Voltage Value