MSA
Appendix: HART Specific Information
ULTIMA X ® Series
116
GB
Command #178: Write Gas Table
Writes the ULTIMA X Gas Table selection. The ULTIMA Gas Monitor uses the Gas
Table value to select a reference table of linearisation values for certain sensors.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
Byte
Format
Description
0
Unsigned
Gas Table selection number [
Table 13.8]
Byte
Format
Description
0
Unsigned
Gas Table selection number
Code
Class
Description
0
Success
No Command-specific errors
1 - 2
Undefined
3
Error
Parameter too large
4
Error
Parameter too small
5
Error
Too few data bytes
6
Undefined
7
Error
In write protect mode
8 - 15
Undefined
16
Error
Access restricted
17 - 31
Undefined
32
Error
Busy
33 - 127
Undefined