Appendix: HART Specific Information
MSA
ULTIMA X ® Series
GB
111
Command #173: Write RTC
Writes the ULTIMA Real Time Clock hours and minutes values.The real time clock
is used to compute the minimum, maximum and average values and to date stamp
the last sensor calibration date.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
Byte
Format
Description
0
Unsigned
RTC hours [0-23]
1
Unsigned
RTC minutes [0-59]
Byte
Format
Description
0
Unsigned
RTC hours [0-23]
1
Unsigned
RTC minutes [0-59]
Code
Class
Description
0
Success
No Command-specific errors
1 - 2
Undefined
3
Error
Parameter too large
4
Undefined
5
Error
Too few data bytes
6
Undefined
7
Error
In write protect mode
8 - 15
Undefined
16
Error
Access Restricted
17 - 31
Undefined
32
Error
Busy
33 - 127
Undefined