MSA
Appendix: HART Specific Information
ULTIMA X ® Series
112
GB
Command #174: Write Alarm Setpoints
Writes the ULTIMA Alarm Setpoint values. The ULTIMA Gas Monitor uses alarm
setpoint values to set alarm status bits in the device. The alarms can be enabled or
disabled, set to increasing or decreasing and can be set to latching. The alarm ad-
justment range is greater than zero and less than full-scale [see "Command 175:
Write Alarm Setpoint Control Actions"].
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
Byte
Format
Description
0
Unsigned
Alarm Number [1, 2, or 3]
1-4
Float
Alarm Setpoint Value
Byte
Format
Description
0
Unsigned
Alarm Number
1-4
Float
Alarm Setpoint Value
Code
Class
Description
0
Success
No Command-specific errors
1 - 2
Undefined
3
Error
Parameter too large
4
Error
Parameter too small
5
Error
Too few data bytes
6
Undefined
7
Error
In write protect mode
8-15
Undefined
16
Error
Access Restricted
17-18
Unrestricted
19
Error
Invalid device variable index
20 - 31
Undefined
32
Error
Busy
33 - 127
Undefined
Содержание ULTIMA X-Series
Страница 1: ...Operating Manual ULTIMA X Series Gas Monitors Order No 10046690 09 ...
Страница 2: ...MSA AUER GmbH Thiemannstrasse 1 D 12059 Berlin Germany MSA AUER GmbH All rights reserved ...
Страница 28: ...MSA Description ULTIMA X Series 28 GB Fig 5 ULTIMA XA Fig 4 ULTIMA Remote sensor module non reactive gas ...
Страница 69: ...Appendix Electrical Installation MSA ULTIMA X Series GB 69 9 3 Installation Outline Drawing CE ULTIMA XA ...
Страница 72: ...MSA Appendix Electrical Installation ULTIMA X Series 72 GB 9 6 HART module M6 109 45 4 309 ...
Страница 158: ...MSA Appendix HART Specific Information ULTIMA X Series 158 GB Sensor Trim Point Screen Fig 52 Sensor trim point screen ...