59
Maintenance and Service
ALTAIR io360
US
(1) Power off the detector.
(2) Press the button for approximately 3 seconds and release.
(3) Remove the housing screws (6 places) with a 5 mm hex key.
(4) Separate the front and back housing.
a) If needed, gently insert the slotted screwdriver into the slots shown.
Do not exceed more than ¼" of penetration.
b) Do not pry where the shield attaches.
c) Rotate screwdriver apart slightly at each appropriate pry location.
(5) Gently pull batteries out of the cavities (3x).