38
Operation
ALTAIR io360
US
6
Operation
6.1
Detector
6.1.1 Interfaces
The detector is operated by a combination of the button, the Calibration Cap, the app, and the Safety io Grid.
6.1.2 Button
The single button on the face of the detector controls the following:
Factory reset will return detector to all factory settings, all network connections to the HUB will be
disconnected, data logs will not be deleted.
Observe proper time for gas reading to stabilize. Response times vary based on the type of
sensor being utilized.
Result
Start State
Action
Powers on
Off
(1) Hold button approximately 3 seconds.
(2) Release.
Powers off
On
(1) Hold button approximately 3 seconds.
(2) Release.
Interactive mode activated:
Bluetooth radio broadcasting*,
Display front light illuminates,
RFID reader activates
On
(1) Press button briefly.
(2) Release.
Initiate Calibration or Bump
On
(1) Press button briefly.
(2) Release.
(3) Present desired RFID tag.
Advance to next gas reading
Interactive mode and no
special states
(1) Press button briefly.
(2) Release.
Acknowledge alarm/alert**
Alarm/alert State
(1) Press button briefly.
(2) Release.
Accept detector configuration Pending configuration
(1) Press button briefly.
(2) Release.
Factory reset
Power up
(1) Hold button approximately 12 seconds.
(2) Release.
Sensor missing accept
Power up
(1) Press button briefly.
(2) Release.
Fault***
On
Button is pressed for approximately 30 seconds
*
When the button is pressed, detector will broadcast Bluetooth for approximately five minute, RFID
communicates for 1 minute.
** If there are multiple gas alarms, the user can acknowledge alarm only when the display shows "Clear".
*** In event the button is stuck pressed, the detector enter a fault state to preserve battery life.
Restart the detector and resume operation.