24
Hardware Setup
ALTAIR io360
US
WARNING!
• Do not install in hazardous location.
• Do not install with the HUB energized.
• Do not touch the circuit board assembly.
• Observe electrostatic procedures.
Failure to follow these warnings can result in serious personal injury or death.
WARNING!
The HUB must be stationary and may be installed on a desk or bench or mounted to a wall while in use.
Follow mounting instructions in the HUB Instruction Manual or damage may occur. Using the product as a
portable device can result in damage that prevents proper operation.
Failure to follow this warning can result in serious personal injury or death.
3.5.1 Power Requirements
(4) Optional:
Use four screws to fasten the wall brackets to the housing.
Recommended M6x16 (DIN EN ISO 4762) fasteners.
(5) Close housing lid by tightening the lid screws with a torque of 13 in-lb
+/- 1 in-lb (1.5 Nm +/- 0.1 Nm).
(6) Close joint.
(7) Install grounding Clamp. Wire cross section for grounding clamp must
be ≥ 0.5mm²
a) Remove grounding clamp.
b) Insert grounding cable and re-tighten grounding clamp with a
torque of 17 in-lb +/- 1 in-lb (2 Nm +/- 0.1 Nm).
Input power requirements:
100 - 240 VAC, 47 - 63 Hz
(Several different prong types are available for world-wide AC sockets)