④
recommend using a dedicated configuration sterilization and loading rack and
basket items.
⑤
Gap should be left between the sterile package, which will help the
penetration factor sterilization.
⑥
large package of difficult sterilization should be placed in the upper , the
smaller packet should be placed on the lower deck.
⑦
sterilizer loading volume shall not exceed 80%.
⑧
only liquid bottles and tubes with the heat load, the load should not exceed the
amount of container volume
7.
equipment daily maintenance,
maintenance
Attention :must disconnect the power before you use the machine and make sure
there is no pressure
To ensure that the sterilizer is in good working condition, and minimize the
number of failures, therefore, the operation described in this chapter must be
followed.
Before starting maintenance, make sure the device is powered down. At the same
time, no pressure in sterile containers.
After the end of the working day with a soft cloth or a piece of gauze to wipe the
door gasket.
Remove the basket.
With gauze with a detergent and water, wipe the inner wall of the sterilization
container. Do not use steel slag wool or steel brush to avoid damaging the
sterilization chamber wall.
Cleaning and sterilizing the container chamber to remove scale. The water in the
tank let go.
Weekly
Once a week, to door sales plus molybdenum disulfide grease.
Once a week, with a soft cloth to wipe the sterilizer housing.
Once a week to clean the filter plug.