he door’s micro-switch loose,
misplace
switch
Under the condition of
warming, Pressure and
temperature do not rise,
or rise slowly
1.Control circuit of the heater
shorts out or burns down
2.pipe joint or safety valve
leaks seriously
1.Check and change the
damaged elements
2.check, tighten the pipe joint,
safety valve, etc
Under the condition of
drainage, Pressure and
temperature do not drop,
or drop slowly
Drain Filter clogs
Clear up the sundries on the
filter valve core
Fail to reach the
sterilization temperature
Whether determined by the
boiling point of the altitude.
Please check the set
temperature of the boiling
point
If not the reason of the
altitude, please contact us or
our distributor
Safety Valve opens
1.Whether the pressure is
too high?
2.Whether the safety valve
breaks down ?
1.Adjust temperature
departure
2. Adjust and change safety
valve
Steam leaking
1.
W
hether door rubber gasket
become hard and aged?
2.
W
hether the door strip cracks?
3.Whether door rubber
gasket falls off?
1. should change door rubber
gasket
2.should change door rubber
gasket
3. re-install door rubber
gasket
2. Alarm Code:
When an error occurs during use, it shows the error code and the buzzer sounds
and indicates alarm, the sterilizer automatically stop moving, please look up the
following conditions and deal with them.
The failure code list
Failure code
Condition of the failure
Failure reason