4
Air valve
Exhaust cold air, increase degree of saturation,
exhaust steam slowly for liquid sterilization
5
water draining solenoid
valve
Drain and exhaust to the built-in circulation tank
after sterilization
6
sterilizing temperature
controller
Drying burning prevention function≤145±5
℃
x
7
dry temperature
controller
Control the temperature of dry heating plate
8
Buzzer
Send the alerting and work end signal
9
Filter
Filter impurities to make the solenoid valve
efficiently
5. Control Panel
There are 9 sterilization programs:
Sterilization
mode/ Program
Times of
pulse
exhaust
The temp.at which
pulse exhaust starts
℃
The temp.at which
pulse exhaust
finishes
℃
Steril
izing
temp
.
℃
Sterilizing time(min)
Drying
time
(min)
Heat
retaining
time(min)
1#Apparatus
1 Sterilizing temp. -2.5
The temp.at which
pulse exhaust starts-
5
1
34
4
5
0
2#Apparatus
bag
2 Sterilizing temp. -2.5
The temp.at which
pulse exhaust starts-
5
1
34
8
10
0
3#Cloth
3 Sterilizing temp. -2.5
The temp.at which
pulse exhaust starts-
5
1
34
12
18
0
4#Rubber
2 Sterilizing temp. -2.5
The temp.at which
pulse exhaust starts-
5
1
21
25
10
0
5#Liquid
0
—
—
1
21
20
0
6#Solid
(Customize)
0-9 Sterilizing temp. -2.5
The temp.at which
pulse exhaust starts-
5
105
~
136
(
1
34
)
(
0-999
)
0
~
99.0
0
7#Liquid
(Customize)
0
—
105
~
136
(
1
34
)
(
0-999
)
0
8#Culture
medium
sterilization
0
-
115
~13
6
0~200
45~60 0~999 min
9#Culture
medium
0
--
0
~80
0~200
45~60 0~999 min