If you want to terminate the process of sterilization , you can press the "Start /
End" button to terminate the sterilization process.
Custom programs can set sterilization temperature and sterilization time:
Choose Program 6# or 7# first, then set as below.
Sterilization temperature settings:
Press the "temperature" button, temperature display window shows the set
temperature and flashes. Through the "
▲
", "
▼
"to turn up and down the
temperature, and then press "temperature" button once to save the temperature
set value.
Sterilization time setting:
Press the "Time" button, the time display window shows the set value and flashes
, by "
▲
", "
▼
" to change the time set value, then press "time" button once to save
the set value.
Parameter setting:
Press the "Start / Stop" and "temperature" button both for 5 seconds to enter the
menu.
Display "iwti" represents the time to add water(unit/minute), when press the
"Start / End"
,
the button shows the value and flashes. Through "
▲
", "
▼
" to
change the value, press "Start / End" to confirm.
Must be 0 displayed "iwti", press the "
▼
", "EATE" displayed, and press the
"
▲
" ,"tErF" displayed . "EATE" represents the cooling end temperature, press the
"Start / Stop", the button displays the value and flashes, through the "
▲
", "
▼
" to
change value, then press the "start / stop" to confirm. The default setting is 100.0
℃
When "EATE" displayed, press the "
▼
", "EATi" showed, press the "
▲
" , "iwti"
showed.
Display "EATi" said the cooling time, in minutes. Press the "Start / End" , the
button shows the value and flashes , through "
▲
", "
▼
" to change value. Then
press "Start / End" to confirm. The default setting is 30 minutes. (After the
cooling, the temperature can't below the set cooling end temperature, otherwise
it will error, if the pot temperature exceeds 105 degrees, the next step will go
directly)