Please make sure that the fixed socket and power plug of power line with same
specification.
Equipment use two phase three wire connection mode, please connect line
according to equipment configuration connection way.
Please do not arbitrarily change the connection mode. If you need, please
contact
us.
Fire wire (L), brown or black, zero line (N) - blue, ground wire (PE) - green and
yellow.
Please entrust a specialized electrical construction personnel to do construction
work.
To ensure your personal safety, please be sure to lay a ground wire.
3. Water source required
Devices do not need to connect the water, you need to add water to sterilizer
water tank or sterilization chamber manually. You are advised to use soft
water
or pure water,
Because if use water which is not suitable may shorten the service life of
equipment, cause unnecessary trouble.Water quality must meet the following
requirements:
Electrical conductivity is less than 15μS/cm
The content of bleach is less than 2mg/L.
PH value is 5~7.
Hardness is less than 0.02mmol/L.
4. Storage Environment
Sterilizer should be stored in a temperature of - 20
℃
~ 55
℃
, relative humidity is
not more than 80%, indoor or sheltered places which is no corrosive gas and
good ventilation.