Err 001
It doesn’t reach the set temperature
after 30 minute’s of pulsation exhaust
air conditioning.
The heater pipe is broken,
have leakage phenomenon.
Err 002
It doesn’t reach the set temperature in
the exhaust air conditioning or the
failure pulsation in duct
The heater pipe is broken.
Improper set parameters
Err 04
During the sterilization, temperature
fluctuates too large
air leaking
Err 08
The sensor improperly connected or
damaged, the temperature in the pot is
below 10%
replace the sensor
Err 016
The failure of circuit board
temperature measuring component.
Replace control board
Err 032
Not add enough water during the set
time of adding water.
Water shortage in the tank,
water float broken, failure
of solenoid valve
Err 064
The water shortage in the pot caused
heating pipe burned
Float failure, after the
heating pipe cool down, add
enough water to the pot
Err 128
After starting sterilization, the pan is
opened
excessive pressure, the pot
shifts, pot test failure
IV. Warranty
To our distinguished guests: Any product has the possibility of failure, please
monitor the work of the equipment in real time during use, If any abnormal,
please first refer to the manual for processing, If the problem still cannot be
solved, you shall inform our Customer Service Centre, to protect against loss.
After-sales service
1. From the day when the equipment is sold(subject to the invoice), one-year
maintenance free of charge and all-life services.
2.The warranty card
:
When you need to normal consultation or maintenance,
please contact to our after-sales service centers with the warranty card and
the
invoice, the warranty card must be well kept.