opening the door by mistake. When water vapor have been released, the
device
will return to the initial position, then the door can be opened.
7.9. Filter Cleaning
Located at the bottom of the device, the filter is used to filter out the impurities,
ensuring pipeline smoothness and electromagnetic valve leak proof. Screw out
filter nut from the chassis of device, remove and clean the filter center. It need
to
be cleaned once a month
approximately
7.10.Cleaning steps of electromagnetic valve
1) remove the cover of the autoclave
2) Pull out the stainless press slip on the electrical magnetic valve with a
screwdriver.
3) lift the electromagnetic valve coil
4) open the valve body with a spanner.
5) clean sundries on the valve center with clean water .
6) re-install the electromagnetic valve.
Third. Common fault and solution
1. This Instruction tries to provide the maintenance method for the known fault,
the below are some common faults.
Phenomena
Possible reasons
Correct Operation
Power switch is on,
power light is not bright.
1.The circuit breaker is
unclosed
2.The main power switch is
damaged
1.Close the circuit breaker
2.Change power switch in
accordance of specific
conditions
The door indicator light
is not bright
1.
T
he door is not fully closed
2.
T
1.Try again after fastening the
door
2.Adjust the door position