Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 17
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
>>Com Baudrate
This sets the baud rate for the RS232 ASCII command set. Available values are: 300, 600, 1200, 2400, 4800,
9600, and 19.2k. Note that the “Special” entry is reserved for future use.
>>Com Line Feed
This controls whether a line feed is added to each carriage return that is sent out by the Model 4215-HS when
the RS232 ASCII command set is being used.
>>Retain Tare
This controls whether the most recent tare value is retained through power down of the unit and used again
immediately on power up. Choosing this option will retain tare on channel A in a single channel unit and both
channels A and B in a twochannel unit.
>>Display Options
This menu contains options that effect how load, peak, and valley are displayed and printed. These are listed in
a submenu that works exactly like the setup mode main menu.
>> Filter Select
There are two types of filters available on the Model 4215-HS: Type I, and Type II. Type I filters are good for
removing most kinds of noise but may leave some jitter on the end digits. Type II are more advanced filters
that are optimized for the typical industrial environment. Both filter types allow you to select from 4 levels of
filtering with Level 1 being the least filtering and Level 4 being the most filtering.
The settling time for each of the 4 levels of filtering is shown in the table below.
Level
1
2
2
4
Setting Time (seconds)
<1
2
10
30
Read Rate (per second)
60
60
30
10