4215-HS Indicator
Manual
(PM-4101)
Sensor View Command (SV)
The Sensor View (SV) command has the format SV. It returns information about the currently selected load cell
including the serial number, the maximum load rating, the millivolt per volt calibration constant, the excitation
voltage used, the latest date of calibration, and the shunt value recorded during that calibration.
Example:
@123SV
Acknowledge:
@123 This is the list of load cell calibration data:
Ch A = S/N 123456, 100.00 Lb , 3.00150 mV/v, 10.00 V , Cal on Oct27-99, 49.532 Lb Shunt
unused S/N 13368, 10.00 Lb , 3.00121 mV/v, 10.00 V, Cal on Oct27-99, 4.9292 Lb Shunt
unused S/N 63220, 100.00 Lb , 2.99984 mV/v, 10.00 V , Cal on Oct27-99, 49.381 Lb Shunt
Ch B = S/N 89991, 1000.0 Lb , 4.50015 mV/v, 5.00 V , Cal on Oct27-99, 486.45 Lb Shunt
Sensor Select Command (SS)
The Sensor Select (SS) command is used to select the calibration data for a stored load cell using the load cell
serial number. The format is SS(channel)(serial number)#.
Example:
@123SSA13368#
Acknowledge
:
@123 This is the list of load cell calibration data:
unused S/N 123456, 100.00 Lb , 3.00150 mV/v, 10.00 V, 49.532 Lb Shunt
Ch A = S/N 13368, 10.00 Lb , 3.00121 mV/v, 10.00 V, 4.9292 Lb Shunt
unused S/N 63220, 100.00 Lb , 2.99984 mV/v, 10.00 V, 49.381 Lb Shunt
Ch B = S/N 89991, 1000.0 Lb , 4.50015 mV/v, 5.00 V, 486.45 Lb Shunt
Sensor Delete Command (SD)
The Sensor Delete (SD) command is used to remove the calibration data for a stored load cell from the list. The
format is SD(serial number)#.
Example:
@123SD13368#
Acknowledge:
@123 Deleted Sensor S/N 13368
Ch A = S/N 123456, 100.00 Lb , 3.00150 mV/v, 10.00 V, 49.532 Lb Shunt
unused S/N 63220, 100.00 Lb , 2.99984 mV/v, 10.00 V, 49.381 Lb Shunt
Ch B = S/N 89991, 1000.0 Lb , 4.50015 mV/v, 5.00 V, 486.45 Lb Shunt