4215-HS Indicator
Manual
(PM-4101)
Example:
@123CMVM0
Acknowledge:
@123 Calibrate Command - Reading for Shunt Check…
[10 seconds delay]
@123 Calibrate Command Completed
Ch A = S/N 123456, 1000.0 Lb , 4.50020 mV/v,
4.50020 mV/v,
4.50020 mV/V,
4.50020 mV/V
4.50020 mV/V
5.00 V , Cal on Oct27-99, 553.26 Lb Shunt
Calibrate by Masses Command (CM)
The Calibrate by Masses (CM) command tells the unit that you are doing a calibration by 2 point or 5 point
mass. The format is CM2 for a two point mass calibration and CM5 for a 5 point mass calibration.
Example:
@123CM5
Acknowledge:
@123 Calibrate by Mass – 5 Point Ready for CMP1 command
Calibrate by Masses Point Command (CMP)
The Calibration by Masses Point (CMP) command can have up to six parts: CMP1 through CMP5 and CMP0. You
must have already started a calibration using the CB and CM commands for any of these commands to func
-
tion. To read the first mass issue the Calibration Mass Point 1 (CMP1) command. The format is CMP1(mass)#.
This will cause the Model 4215-HS to read 10 seconds of data from the load cell. These readings are averaged
to obtain the first calibration mass reading. The example uses 0.0 Lb for the first mass. Be sure you have placed
the mass on the load cell before entering this command since reading will begin immediately.
Example:
@123CMP10.0#
Acknowledge:
@123 Calibrate Mass 1 Command - Reading...
[10 seconds delay]
Calibrate Mass 1 Command - Ready for CMP2 or CE command
The Calibrate Mass 2 (CMP2) command has the format CMP2(mass)#. This command can only be issued after
the CMP1 command has completed. The example uses 500 Lb for the second calibration mass. Be sure you
have placed the mass on the load cell before entering this command since reading will begin immediately.