4215-HS Indicator
Manual
(PM-4101)
* Print current readings of load, peak, valley, position, and velocity using the currently selected units at the
touch of one button.
* Automatic identification of calibrated load cells with auto-id option.
* Calibration for load cells using manufacturer’s milli-volt per volt calibration factor, 5 point milli-volt per volt,
2 or 5 point known mass, or internal precision shunt calibration. A back panel switch allows you to choose be
-
tween two values used for the shunt calibration so there is no need to open the unit.
* Two Channels of Analog output of any signal (load, peak, valley, position, velocity) with user defined scale
factor and offset. Large, buffered ± 10.00 volt output.
* Remote operation using standard RS232 ASCII command set.
* Back panel digital inputs for tare, peak and valley reset, position reset, and print. Other functions available as
custom programmed options.
* Optionally measure the average force using a contact closure to begin and end the averaging period.
This section will help you get your Model 4215-HS set up and operating in just a couple of minutes. The back
panel is shown below. You will need to make at least two connections here. First be sure that the power switch
is off (the lower part of the rocker should be pressed in). Connect the AC power line and plug it in.
Next connect a load cell to the 9 pin female connector labeled “Sensor A”. You should use the cabling supplied
with your Model 4215-HS and load cell. If you do not have a factory supplied cable, see Appendix B for the
recommended load cell connection.
Model 4215-HS Back Panel
Quick Start