Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 12
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
>Analog Output Channels
This menu is used to set up the analog outputs. Instead of a submenu there is a series of selections. Pressing
escape at any time will leave the analog output unchanged and return you to the setup mode main menu.
1. Channel: -- This entry allows you to select the analog output channel A or B. (use either the left/right or
plus/minus buttons) for the selected source data that will drive the analog output. Press enter to select the
displayed Analog channel.
2. Source: -- This entry allows you to scroll through a list (using the left/right or plus/minus buttons) to select
the source of the data that will drive the analog output. The standard list will include load, peak, valley, posi
-
tion, and velocity. Press enter to select the displayed item.
3. Units: -- This entry allows you to scroll through the list of available units for the data source you have chosen
in the previous step.
4. Full -- This entry specifies the value of source data (in the selected units) that will cause the Model 4215-HS
to output positive full scale analog voltage (+10.0 volts).
5. Zero -- This entry specifies the value of source data (in the selected units) that will cause the Model 4215-HS
to output zero analog voltage (0.0 volts).
Once you have entered these 4 selections the analog output is setup.
>Sensor Select
This menu is used to manually select load cell calibration data from a list stored in the Model 4215-HS. When
you select this menu on a 2 channel unit, you are asked to select Channel A or B. The load cell serial number of
the selected channel is displayed. Use either the left/right or plus/minus buttons to page through the list. The
escape button will return to the setup mode main menu without changing the load cell that is selected. The
enter button will select the currently displayed load cell and briefly display the rated load, last calibration date,
and excitation voltage.
To add new load cells to this list use the Calibration entry on the setup mode main menu described below.
If you are using auto-id load cells this menu will allow you to see which load cell is selected but will not allow
you to change the selection. This ensures that the correct calibration informationis always used when you use
Auto-ID load cells.