V304
is lit:
The current limit for the load cell supply is activated.
Causes:
-
There is a short circuit in the load cell supply lines
-
The load cell resistance is too low (e.g. too many load cells connected)
Current limiting at >96 mA
WARNING
Working on the device while it is switched on may have life-threatening
consequences.
When removing covers or parts using tools, live parts or terminals may be exposed.
Please note that capacitors in the device may still be charged even after disconnecting
the device from all voltage sources.
Disconnecting the device from the power supply.
Disconnect the device from the power supply and eliminate the fault.
After the fault is eliminated it may be necessary to switch off the device and/or
disconnect the load cells in order to deactivate current limiting.
16.8 Error numbers @ "LAST_ERROR"
16.8.1
Weighing point error
Number
Display
Cause
2
in use
Weighing point is in use, e. g., by an application.
6
test active
Test is active, no weights.
7
cali active
Calibration is active, no weights.
8
no standstill
No standstill of the scale.
13
tare is active
Tare is active.
15
weight is dimmed
Weight is not legal-for-trade (<0 or >max.).
16
weight has error
Weight error.
17
scale not ready
Scale is not ready.
18
cannot tare below zero
Taring below zero is not possible.
102
timeout
Only on xBPI scales: Time limit exceeded when sending a com-
mand to the scale.
142
calibration active
During calibration, taring and zeroing is not possible.
147
no zeroset
Zeroset outside of the zero setting range is not possible.
149
Busy
The scale is currently busy with another query.
255
hardware error
Weighing point is faulty.
Transmitter in field housing PR 5230