![Minebea Intec PR 5230 Скачать руководство пользователя страница 271](http://html1.mh-extra.com/html/minebea-intec/pr-5230/pr-5230_instrument-manual_1801221271.webp)
12.2.6.11
Register 30, 31: Fixed values (read/write)
Register 30
Fixed value for analog output; value (num) 0
20000 corresponds to
20 mA
Register 31
Fixed value for FixTare, see also
SetFixTare
,
GetFixTare
in Chap-
ter
.
12.2.6.12
Register 80
89: State-controlled action bits (write)
Note:
For setting bits, see Chapter
.
Only setting and resetting of single bits is possible.
When changing a bit from 0 to 1, the corresponding action starts. After handling the
command, the bit must be reset. Application: The master writes cyclically.
The bit is set as
Write_Value_Select
with the specified number (see Chapter
).
The bit is reset at the specified 128.
Register 80
SetZero
Set the gross weight to zero.
Register 81
SetTare
Execute taring.
Register 82
ResetTare
Reset tare.
Register 83
SetTest
Start the ADC test.
Register 84
ResetTest
Finish the ADC test.
Register 85
ResetPwf
Reset the PowerFail bit (Register 1; the bit was
set after power on).
Register 86
SetFixTare
Taring with weight in numerical address D 31
"Fixtare."
Register 87
GetFixTare
The current gross weight is copied into numeri-
cal address D31.
Register 89
ResetError
The
CmdError
error bit is reset.
12.2.6.13
Register 112
121: Transition-controlled action bits (write)
For setting bits, see Chapter
As soon as the bit was set, it is reset internally and the operation is handled; this is
transition-controlled (for writing once).
The bit is set as
Write_Value_Select
with the specified number (see Chapter
).
Register 112
SetZero
Register 113
SetTare
Register 114
ResetTare
Register 115
SetTest
Register 116
ResetTest
Transmitter in field housing PR 5230
12 Fieldbus interface
Minebea Intec
EN-268