Function
SPM Bit
Description
marker bit 2
X65 = 0
Marker bit 2 not set, after power-on the
markers are set to "0".
marker bit 3
X66 = 0
Marker bit 3 not set, after power-on the
markers are set to "0".
actual diginp1
X00 = 1
digital input 1: active
actual diginp2
X01 = 1
digital input 2: active
actual diginp3
X02 = 1
digital input 3: active
limit 1 out
X16 = 1
Limiting signal 1: active
limit 2 out
X17 = 1
Limiting signal 2: active
limit 3 out
X18 = 1
Limiting signal 3: active
ADC error
X32 = 1
General error in the weighing point
above Max
X33 = 1
Weight above Max
overload
X34 = 1
Weight above Max plus the 'overload'
value
below zero
X35 = 1
Weight below zero
center zero
X36 = 1
Weight within ¼ d of zero
inside ZSR
X37 = 1
Weight within zero-setting range
standstill
X38 = 1
Standstill is active
out
X39 = 1
Weight below zero or above Max
command error
X48 = 1
For internal use only.
command busy
X49 = 1
For internal use only.
power fail
X50 = 1
Set after power-on (=power failure)
test active
X56 = 1
Analog test was started.
cal active
X57 = 1
For internal use only.
tare active
X58 = 1
Instrument is tared.
marker bit 1
X64 = 1
Marker bit 1 set, after power-on the
markers are set to "0".
marker bit 2
X65 = 1
Marker bit 2 set, after power-on the
markers are set to "0".
marker bit 3
X66 = 1
Marker bit 3 set, after power-on the
markers are set to "0".
Example: [overload] function
SPM Bit [X34 = 1]
Function and output are active (e.g.: if 'overload' is reached, a lamp is lit).
SPM Bit [X34 = 0]
Function is active and output is not active (e.g.: if "overload" is reached, a lamp goes out).
The parameters for the digital outputs are defined under
-[Digital i/o parameter].
7 Commissioning
Transmitter in field housing PR 5230
EN-205
Minebea Intec