6-11
In P-A/C under NIV, you need to set the following ventilation parameters:
1. [
O2%
]:
Oxygen
concentration
2. [
Pinsp
]:
Pressure
control
level
of
inspiration
3. [
Tinsp
] or [
I:E
]:
Time of inspiration or ratio between the inspiratory and
expiratory time
4. [
f
]:
Breathing
frequency
5. [
Tslope
]:
Time
of
pressure
rising
6 [
PEEP
]:
Positive
end-expiratory
pressure
7. [
Assist
]:
Assisted
trigger
8. [
F-Trig
] or [
P-Trig
]: Inspiration
trigger
level
6.7.6 CPAP/PSV
PSV, pressure support ventilation, is a pressure mode. The system delivers a PSV when it
detects that patient inspiratory effort reaches the preset inspiration trigger level. Time of
pressure rising and pressure support level are set by the user. At the beginning of inspiratory
phase, patient airway pressure rises to the preset pressure level within the preset Time of
pressure rising and is held at this pressure level until patient inspiratory flow is detected to
reach the expiration trigger level. In PSV, when the airway pressure is held at the preset
pressure level, delivered gas flow changes with the resistance and compliance of the patient’s
lungs.
Paw
Flow
Time
Apnea time
△
Papnea
Apnea ventilation
Apnea vent cycle
I:E=1:2
Tslope
△
Psupp
PSV
Exp. trigger
Insp. trigger
Insp. trigger
Содержание SynoVent E5
Страница 1: ...SynoVent E5 Ventilator Operator s Manual...
Страница 2: ......
Страница 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Страница 46: ...3 18 FOR YOUR NOTES...
Страница 60: ...4 14 FOR YOUR NOTES...
Страница 68: ...5 8 FOR YOUR NOTES...
Страница 92: ...6 24 FOR YOUR NOTES...
Страница 118: ...9 12 FOR YOUR NOTES...
Страница 154: ...A 8 FOR YOUR NOTES...
Страница 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Страница 170: ...B 16 FOR YOUR NOTES...
Страница 184: ...D 8 FOR YOUR NOTES...
Страница 195: ......
Страница 196: ...P N 046 000991 00 10 0...