E-1
E
Factory Defaults
This chapter lists the most important factory default settings which are not user-adjustable.
When necessary, you can restore the factory default settings.
E.1 CO2 Module
CO2 module
Factory default setting
Working mode
Measure
Pump rate
100 mL/min
Humidity compensation
OFF
CO2 calibration concentration
setting
3%
EtCO2 low limit
Adult: 15 mmHg; Pediatric: 20 mmHg
EtCO2 high limit
Adult/pediatric: 50 mmHg
FiCO2 high limit
Adult/pediatric: 4 mmHg
E.2 Alarm
Alarm
Factory default setting
Paw high limit
50 cmH2O
MV high limit
TV*f*1.5
MV low limit
TV*f*0.8
TVe high limit
TV setting value x 2
ftot high limit
OFF
Tapnea 15s
Alarm Volume
2