4-7
4.3 Spirometry Loops Screen
Select [
Spirometry
] to access the spirometry loops screen as shown below.
Spirometry loops reflect patient lungs function and ventilation condition as well, such as the
patient’s lungs compliance, over-inflation, breathing system leakage and airway blockage.
The system provides three types of spirometry loops: P-V (pressure-volume) loop, F-V
(flow-volume) loop, and F-P (flow-pressure) loop. The three types of loops come from
pressure, flow, and volume waveforms data.
Up to two types of spirometry loops are displayed at a time. To select the desired loop:
1. Select
[
Spirometry
] on the main screen.
2. Select the desired loop to be displayed.
The ventilator provides the function of reference loop. Selecting [
Save
] saves the current F-V
loop, P-V loop, and F-P loop as reference loop and displays the time on which the reference
loop is saved. Selecting the time button views the reference loop saved at that time moment.
Selecting [
Hide Ref.
] hides the reference loop which is being displayed.
The ventilator saves reference loops at up to four time moments. If reference loops at four
Содержание SynoVent E5
Страница 1: ...SynoVent E5 Ventilator Operator s Manual...
Страница 2: ......
Страница 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Страница 46: ...3 18 FOR YOUR NOTES...
Страница 60: ...4 14 FOR YOUR NOTES...
Страница 68: ...5 8 FOR YOUR NOTES...
Страница 92: ...6 24 FOR YOUR NOTES...
Страница 118: ...9 12 FOR YOUR NOTES...
Страница 154: ...A 8 FOR YOUR NOTES...
Страница 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Страница 170: ...B 16 FOR YOUR NOTES...
Страница 184: ...D 8 FOR YOUR NOTES...
Страница 195: ......
Страница 196: ...P N 046 000991 00 10 0...