3-8
The ventilator works normally under supply pressure of 280 to 650 kPa. Supply pressure of
less than 280 kPa can impair the performance of the ventilator and even cease ventilation.
Supply pressure between 650 and 1000 kPa can impair the performance of the ventilator but
will not cause any hazard arising from high pressure gas.
3.4 Install the Support Arm
A.
Fixing
block
knob
B.
Fixing
block
C.
Hose
hook
D.
Support
bar
E.
Support
arm joint
F. Support arm joint
G. Support arm joint
1. Loosen the fixing block knob. Place the fixing block onto the handle on the side of the
ventilator.
2. Tighten the fixing block knob.
D
C
B
A
F
E
G