B-10
B.9 Special Functions
Function
Specification
Inspiration
Hold
Push and hold the Insp. Hold key to activate this function.
Inspiration Hold is active for a maximum of 30s.
Expiration
Hold
Push and hold the Exp. Hold key to activate this function.
Expiration Hold is active for a maximum of 30s.
O2
↑
O2
↑
is delivered for a fixed 2 min.
During O2
↑
, O2 concentration for adult patients is 100% and that for pediatric
patients is 1.25 times of the currently set O2 concentration or 100%, whichever is
less.
Suction Phase
1:
O2
↑
before suction. Delivering 100% O2 lasts for a maximum of 120 s. O2
concentration for adult patients is 100% and that for pediatric patients is 1.25 times
of the currently set O2 concentration or 100%, whichever is less. When patient
disconnection is detected, the system enters next phase automatically.
Phase 2: suction. Suction lasts for a maximum of 120s. When patient reconnection is
detected, the system enters next phase automatically.
Phase 3: O2
↑
after suction. Delivering 100% O2 lasts for a maximum of 120s. O2
concentration for adult patients is 100% and that for pediatric patients is 1.25 times
of the currently set O2 concentration or 100%, whichever is less.
Nebulizer Supports
jet
nebulizer;
Supports to set nebulizer time ranging from 1 to 60 min.
Manual
Breath
One breath is delivered in the expiratory stage.
Manual breath is not responded if delivered in the inspiratory stage or when the
expiratory stage is not finished.
Sigh
Sigh is started once every three minutes in V-A/C. Sigh started within two
continuous ventilation cycles is effective.
Volume
sigh
A sigh is delivered every 100 breaths or 7 minutes (whichever is earlier) in V-A/C
mode. Tidal volume of 1.5 times of the set value is delivered every 100 breaths.
Screen
Locking
Prevents ventilator settings and values displayed from being changed due to
inadvertent key clicking.
Содержание SynoVent E5
Страница 1: ...SynoVent E5 Ventilator Operator s Manual...
Страница 2: ......
Страница 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Страница 46: ...3 18 FOR YOUR NOTES...
Страница 60: ...4 14 FOR YOUR NOTES...
Страница 68: ...5 8 FOR YOUR NOTES...
Страница 92: ...6 24 FOR YOUR NOTES...
Страница 118: ...9 12 FOR YOUR NOTES...
Страница 154: ...A 8 FOR YOUR NOTES...
Страница 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Страница 170: ...B 16 FOR YOUR NOTES...
Страница 184: ...D 8 FOR YOUR NOTES...
Страница 195: ......
Страница 196: ...P N 046 000991 00 10 0...