6-12
In CPAP (continuous positive airway pressure ventilation), the airway pressure is held at the
user-set positive pressure level throughout the ventilation cycle. The patient breathes
spontaneously and determines his own breathing frequency, tidal volume, and breath time.
The system starts apnea ventilation when detecting that the period of time when patient does
not perform continuous spontaneous breathing exceeds the preset apnea time.
Paw
Flow
Time
△
Papnea
Spontaneous breathing
Apnea ventilation
PEEP
Apnea vent cycle
I:E=1:2
Tslope
Apnea time
Time
In CPAP/PSV under non-NIV, you need to set the following ventilation parameters:
1. [
O2%
]:
Oxygen
concentration
2. [
Tslope
]:
Time
of
pressure
rising
3. [
△
Psupp
]:
Pressure support level
4 [
PEEP
]:
Positive
end-expiratory
pressure
5.
[
Exp
%
]:
Expiration
trigger
level
6. [
Δ
Papnea
]:
Pressure
of
apnea
ventilation
7. [
fapnea
]:
Frequency
of
apnea
ventilation
8. [
F-Trig
] or [
P-Trig
]: Inspiration
trigger
level
9. [
ATRC
]: Automatic tube resistance compensation
In CPAP/PSV under NIV, you need to set the following ventilation parameters:
1. [
O2%
]:
Oxygen
concentration
2. [
Tslope
]:
Time
of
pressure
rising
3. [
△
Psupp
]:
Pressure support level
4 [
PEEP
]:
Positive
end-expiratory
pressure
Содержание SynoVent E5
Страница 1: ...SynoVent E5 Ventilator Operator s Manual...
Страница 2: ......
Страница 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Страница 46: ...3 18 FOR YOUR NOTES...
Страница 60: ...4 14 FOR YOUR NOTES...
Страница 68: ...5 8 FOR YOUR NOTES...
Страница 92: ...6 24 FOR YOUR NOTES...
Страница 118: ...9 12 FOR YOUR NOTES...
Страница 154: ...A 8 FOR YOUR NOTES...
Страница 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Страница 170: ...B 16 FOR YOUR NOTES...
Страница 184: ...D 8 FOR YOUR NOTES...
Страница 195: ......
Страница 196: ...P N 046 000991 00 10 0...