Ventilator Control and Drive
Theory of Operation
1 - 24
Sensor Board
Pair, J1
PIN NAME
FUNCTION
1. AGND Analog
Ground
2.
Pair
Pressure of air
3. NC
No
connection
4. AR10V 10V
5. NC
No
connection
PAW, J2
PIN NAME FUNCTION
1. AGND Analog
Ground
2.
Paw
Pressure inside airway
3. AR5V 5V
4. NC
No
connection
5. NC
No
connection
Work Light Power, J11
PIN NAME FUNCTION
1. 24V
Power
24V
2. GND Power
Ground
Backlight Power, J15
PIN NAME FUNCTION
1. 24V
Power
24V
2. GND Power
Ground
Switch of Gas Supply, J6
PIN NAME
FUNCTION
1. GND Ground
2.
NC No
connection
3. Pgas Switch
input
Expire Valve, J14
PIN NAME FUNCTION
1. 5V
5V
2.
Valexp Expire
Valve Drive