Calibration Procedures
Calibration
5 - 12
5.5.7
Flow Meter
1. Calibration
FIGURE 5-9
Connecting the Flow Analyzer
a.
Connect one end of the Respiration Tube to the Common Gas Outlet on the
, and the other to the inlet on the Flow Analyzer (Certifier-FA PLUS).
b.
Set the Fresh Gas flow to 0 mL/min.
c.
Select
Continue
to start calibration.
d.
Ensure that there is 0 flow on the Flow Analyzer.
e.
Set the on screen calibration to
0
, and select
Done
to begin calibration. Successful
calibration will be confirmed by illuminating the corresponding calibration value
yellow. If the calibration value illuminates red, recalibrate the value until it
successfully calibrated.
f.
Set the onscreen calibration value to
1
.
g.
Raise the flow until it reads 1.00 on the Flow Analyzer using the Flow Meter.
h.
Select
Done
to begin calibration.
i.
Repeat steps
f.
through
h.
for all calibration values listed on the screen. Ensure all
successfully calibrate and illuminate yellow.
j.
After all values are successfully calibrated select
Finish Cal
.
k.
The
AS3000
will confirm that calibration was successfully completed.
l.
Select
Done
to return to the calibration screen.
NOTE:
When calibrating the Flow Meter, use AIR if available,
otherwise use a combination of O
2
and N
2
O to attain a flow
of 15 L/min.
Содержание Datascope AS 3000
Страница 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Страница 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 127: ...Calibration Procedures Calibration 5 14 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 178: ...AS3000 Service Manual 0070 10 0683 7 3 Warranty Manufacturer s Responsibility This page intentionally left blank...
Страница 179: ...0070 10 0683 Rev K August 8 2014...