Installation Checkout Procedure
Installation Guide
2 - 26
n.
For setting
0
and
T
there should be no output of anesthetic agent.
• At 1 vol.%; 0.8 - 1.2 vol.% *
• At 2 vol.%; 1.8 - 2.2 vol.% *
• At 3 vol.%; 2.8 - 3.2 vol.% *
• At 4 vol.%; 3.8 - 4.2 vol.% *
• At 5 vol.%; 4.8 - 5.2 vol.% *
• At 6 vol.%; 5.7 - 6.3 vol.% *
• At 7 vol.%; 6.7 - 7.3 vol.% *
• At 8 vol.%; 7.7 - 8.3 vol.% *
* = Correct for temperature and carrier gas if necessary.
o.
Switch off the vaporizer until
0
engages.
p.
Switch off the AIR or O
2
flow.