6 - 11
Periodic Maintenance
Check Valve Cleaning
7. Install absorber canister assembly:
1.
Apply Loctite 243 to (3) screws of the Absorber Mounting Plate.
2.
Secure mounting plate to Breathing System Block using (3) screws and (3) associated
washers.
3.
Apply Loctite 243 to the threads of (2) Absorber Guide Rods.
4.
Install rods into block until rod height is 314 +/- 0.5 mm (old style breathing systems)
or 303+/-0.25 mm (new style breathing systems) (see FIGURE 6-9).
FIGURE 6-8
Absorber Canister Assembly
FIGURE 6-9
Breathing System Styles
Absorber
Mounting Plate
Rod, Support
Absorber
Canister (2 pcs)
Holder,
Absorber
Canister
Gasket,
Absorber
Canister
Gasket,
Absorber
Ring Seal
(inside
Absorber
Canister)
New Style Breathing Systems
(115-005967-00)
Old Style Breathing Systems
(0997-00-0618)