AS3
0070-10-0683
6 - 3
Perio
d
ic Mainte
nance
Replace
m
ent of Co
nsuma
ble Par
ts
6.5
Replacement of Consumable Parts
Parts are replaced at multiple intervals from the date of installation.
CONSUMABLE PARTS
AMOUNT
12 MONTH PM
36 MONTH PM
PART NUMBER
1
O
2
Sensor
1
check
check
0600-00-0149
2
Battery Assembly SLA-5AH24V
1
check
replace
0997-00-0619
3
Drive Gas Tube
1
check
replace
0004-00-0098
4
Bellows Assembly
1
replace
replace
115-005968-00
5
Scavenger Hose
1
check
replace
0004-00-0095
6
I/E Valve gasket
2
replace
replace
0354-00-0192
7
O-ring, 6 x 1.8 (for airway pressure gauge)
1
replace
replace
0354-00-0188
8
O-ring, 4 x 1.8 (for airway pressure gauge)
1
replace
replace
0354-00-0189
9
Bag Arm retainer ring
1
check
check
0354-00-0039
10
Disc, Ceramic I/E Valve
2
check
replace
0354-00-0182
11
Gasket, Bellows canister base
1
replace
replace
0354-00-0191
12
Absorber canister Seal
3
check
replace
0354-00-0183 (original)
1
2
check
check
replace
replace
047-002850-00 (revised)
047-000252-00 (revised)
13
O-ring O
2
Interface / O
2
Plug
2
replace
replace
0354-00-0187
14
O-ring Vaporizer mount 14x2.65
4
replace
replace
0354-00-0193
15
Cylinder Washer, Black
3
replace
replace
0348-00-0185
16
Absorber assembly
1
clean
clean
0997-00-0629 (original)
115-005724-00 (revised)
17
AGSS filter
1
clean
replace
0378-00-0074
18
Absorber Gasket
2
clean
replace
0354-00-0199 (original)
047-002849-00 (revised)
19
Check Valve
(Applicable to original model P/N 0997-00-0618
only. See “Check Valve Cleaning” on page 6-4.)
1
clean
clean
Non-replaceable
20
User Interface cable jackscrews and jackposts
2
Check for Loctite
243 and apply if
necessary. Verify
jackscrews are
secured tightly.
Check for Loctite
243 and apply if
necessary. Verify
jackscrews are
secured tightly.
Non-replaceable
Содержание Datascope AS 3000
Страница 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Страница 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 127: ...Calibration Procedures Calibration 5 14 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 178: ...AS3000 Service Manual 0070 10 0683 7 3 Warranty Manufacturer s Responsibility This page intentionally left blank...
Страница 179: ...0070 10 0683 Rev K August 8 2014...