Installation Checkout Procedure
Installation Guide
2 - 16
c.
Select
Continue
to prompt to the Normal Screen.
14. Manual Leak Test
NOTE:
The Manual Leak Test detects smaller leaks than can be
detected in the Automatic Leak Test.
a.
Ensure that the gas pressure for O
2
, N
2
O, and AIR are at 50 ±10 psi.
b.
Power ON the
.
c.
Attach a reusable-silicone-rubber breathing circuit to the Breathing System.
NOTE:
For testing purposes always use a reusable breathing
circuit.
d.
Tightly connect the Y-fitting on the breathing circuit to the test port.
e.
Attach a breathing bag to the bag arm.
f.
Set the APL Valve to the fully closed position (
70
cmH
2
O).
g.
Rotate the O
2
Flow Control Valve until 50 cmH
2
O pressure is observed on the
Airway Pressure Gauge.
h.
Verify that the flow required to stabilize the pressure is less than 300 mL/min.
FIGURE 2-12
Compliance Test Results