2 - 21
Installation Guide
Installation Checkout Procedure
c.
Attach a Vent Tester between the EXP port and the expiratory hose.
d.
Set the O
2
flow to 2 L/min and set the N
2
O and AIR flow rates to minimum flow.
e.
Set the ventilator controls to:
f.
Select
CMV
to begin ventilation.
g.
Verify that the pressure waveform, Tidal Volume, Mean or Plateau Pressure, Resp.
rate and minute volume values appear on the screen.
h.
Verify the Tidal Volume display is within 25 ml or ±15% of the delivered volume
(whichever is greater) measured within approximately 1 minute from the start of
ventilation.
i.
Verify the delivered volume as measured by a Vent Tester at the expiratory port, is
within 17% of the Tidal Volume set value within approximately 1 minute from the
start of ventilation.
j.
Verify that the PEEP reading is 4 ±2 within approximately 1 minute from the start of
ventilation.
20. Verify that the PCV ADULT ventilator mode operates.
a.
Attach a breathing circuit and breathing bag.
NOTE:
For testing purposes always use a reusable breathing
circuit.
b.
Attach an adult test lung to the Y-fitting of the breathing circuit.
c.
Attach a Vent Tester between the EXP port and the expiratory hose.
d.
Set the O
2
flow to 3 L/min and set the N
2
O and AIR flow rates to minimum flow.
VENTILATOR CONTROLS
VENTILATOR SETTINGS
Patient Type
Child
Ventilation Mode
CMV
Tidal Volume -
V
T
120
Breath Rate -
freq
20
I:E Ratio -
I:E
1:2
Plateau -
T
P
10
PEEP -
PEEP
Off
Содержание Datascope AS 3000
Страница 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Страница 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 127: ...Calibration Procedures Calibration 5 14 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 178: ...AS3000 Service Manual 0070 10 0683 7 3 Warranty Manufacturer s Responsibility This page intentionally left blank...
Страница 179: ...0070 10 0683 Rev K August 8 2014...