B-1
B
EMC and Radio Regulatory Compliance
B.1 EMC
The device meets the requirements of IEC 60601-1-2:2014.
NOTE
Use of accessories, transducers and cables other than those specified
or provided by the manufacturer of this device could result in increased
electromagnetic emissions or decreased electromagnetic immunity of
this device and result in improper operation.
Use of this device adjacent to or stacked with other device should be
avoided because it could result in improper operation. If such use is
necessary, this device and the other device should be observed to verify
that they are operating normally.
Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30
cm (12 inches) to any part of the this device, including cables specified
by the manufacturer. Otherwise, degradation of the performance of this
device could result.
The non-ME EQUIPMENT (e.g. ITE) that is a part of an ME SYSTEM may
be disrupted by the electromagnetic interference of nearby equipment. It
may be necessary to take mitigation measures, such as re-orienting or
relocating the non-ME EQUIPMENT or shielding the location.
This device is intended for use in professional healthcare facility
environment only. If it is used in special environment, such as magnetic
resonance imaging environment, the equipment/system may be
disrupted by the operation of nearby equipment.